User's Manual

Table Of Contents
Table of contents
18
Set an immediate bolus (Variable Bolus option is “on”)
1. From the Time/Date screen press SEL until
the “BOLUS” screen is displayed. The last
bolus value programmed and the Time and
Date will be flashing.
2. Press ACT and “SET BOLUS TYPE”
appears. Use the and buttons to choose
“NORMAL.”
3. Press ACT and “IMM” appears with the
dashes underneath flashing.
4. Use the and buttons to enter an
immediate bolus
amount
5. Press ACT and the screen shows “CONFIRM
IMM.”
6. If the number was not correct, wait for the
screen to return to Time/Date and then repeat
this procedure to enter the correct value.
7. If the number you entered for the immediate
bolus is correct, press ACT to confirm it.
Place the PPC near the Pump to complete the
programming.
8. The PPC will beep once and display the
amount of insulin delivered. If the Alarm
Feedback feature is “ON,” the Pump will
beep during each of the first five strokes.
9. The PPC will beep when the bolus delivery is
complete. Three segments of the insulin
delivery icon will be displayed and spinning slowly during the bolus
delivery. By pressing SEL you can see the amount of insulin being
delivered.
08:23 Jan 04
BOLUS
_ _
IMM EXT
_ _U
_ _U
SET
NORMAL
BOLUS TYPE
SET BOLUS
IMM
_ _U
CONFIRM
IMM
2.6U
PPC
COMMUNICATING
08:23 JAN 04
BOLUS 0.0U