User's Manual

Table Of Contents
Table of contents
31
Auto off
“AUTO OFF” sets an alarm on your PPC to remind you if you have not pro-
grammed your Pump recently. The “AUTO OFF” alarm can be set between 1
and 16 hours. To turn off the "AUTO OFF" program set the time to dashes.
1. From the “SETUP PUMP” screen, press ACT
then SEL.
2. The screen will display “AUTO OFF” with
flashing dashes. Press ACT.
3. Use the and buttons to select the number
of hours before an Auto Off alarm occurs.
4. Press ACT. In this example, a time duration of
10 hours was selected. The PPC will alarm if
you did not program your Pump during the
past 10 hours.
5. Place the PPC near the Pump. The PPC will
beep once when the communication is
completed.
Alarms
There are three PPC Alarm options, two audible tones (Low/High) and a
vibrate mode.“ALARMS” will alert you if either the PPC or Pump recognizes
a problem. The vibrate only alarm provides all programming feedback from
PPC with beeps.
1. Press ACT on the ALARMS menu to enter
the “ALARMS” menu.
AUTO OFF
_ _ HRS
AUTO OFF
10 HRS
SET
PPC
COMMUNICATING
ALARMS