2. Press the and buttons to select the type
of alarm you want, then press ACT.
3. The screen will now display “SET
ALARM FEEDBACK.” This setting should
always be “ON”. Press ACT.
4. Place the PPC near the Pump. When the
communication is completed, the PPC will
change to “SELF TEST” and then timeout to
the Time / Day screen.
Self test
Your doctor may ask you to run a diagnostic test of your Pump System.
“SELF TEST” will send messages between your PPC and Pump to check
their operation.
NOTE: Notify your doctor if any of these Self Test events do
not occur.
If the PPC displays a Medtronic MiniMed logo and
software number, the PPC has re-started but has a low
battery. Replace the battery immediately.
1. From the “SETUP PUMP” screen, press SEL
until the “SELF TEST” screen is displayed.
2. Press ACT.
3. Place the PPC near the Pump and complete
the communication process.
4. Verify the following events occur:
The Pump will beep four times.
The PPC backlight will turn on.
SET
PPC
ALARM TYPE
LOW/HIGH/VIBRATE
SET
ALARM
FEEDBACK
ON/OFF
PPC
COMMUNICATING
SELF TEST
PPC
COMMUNICATING