User's Manual

Table Of Contents
Table of contents
36
Use audio bolus in the Main Menu
NOTE: When an immediate bolus is being delivered, the
button will be disabled.
From the Time/Date screen press button. Each time the button is pressed
the PPC will beep one or two times, depending upon the step size set (0.4U or
0.8U).
1. Count the number of beeps to determine how
much insulin you want to program. The bolus
amount delivered is determined by the step
size and the number of times the button is
pressed. For example: If the step size is
programmed to 0.4U and the button is
pressed three times, the total bolus delivered will be 1.2U (3 x 0.4U).
2. Press ACT, and the beeps repeat to confirm
the bolus amount you have entered.
3. Press ACT twice, and the PPC will program
the Pump. The PPC will beep once when the
communication is completed.
Turn on the Variable Bolus feature
If “OFF” is chosen, the Variable Bolus feature will not be available.
1. From the “SETUP II” menu, press SEL until
the “VARIABLE BOLUS” screen is
displayed.
2. Press ACT and the screen changes to “SET
VARIABLE BOLUS.” The word “ON” is
flashing.
3. Use the and buttons to select “ON” or
“OFF.”
SET BOLUS
IMM
12U
PPC
COMMUNICATING
VARIABLE ON/OFF
BOLUS
VARIABLE ON
BOLUS