If the PPC main battery (AA 1.5V alkaline) has been replaced or the PPC rec-
ognizes the “PUMP STATUS” needs to be checked, the following message is
displayed:
Place the PPC near the pump, then press SEL and ACT. Allow the communi-
cation to complete.
NOTE: If while programming the Pump, the PPC goes blank,
the PPC beeps six times and then the "CHECK PUMP
STATUS" message appears, the PPC (AA 1.5V
alkaline) battery needs to be replaced. For
instructions, see "Installing/Replacing the Main
Battery" in Chapter 8.
Pump suspended
If the Pump operation has been suspended, the following message is dis-
played:
During “SUSPEND PUMP,” the Pump will deliver a basal rate of
0.2 U/h. To restart insulin delivery programming, press SEL and ACT. Then
place the PPC near the Pump and allow the communication to complete.
PPC display Type of alarm
CHECK PUMP STATUS 1
PPC display Type of alarm
PUMP
SUSPENDED
2