Personal Pump Communicator messages
Display screen message Message meaning or action required
AUTO OFF PUMP
PUMP SUSPENDED
Auto Off time interval has elapsed. Pump operation is suspended.
BOLUS 0.0 u A bolus has been programmed and is being delivered.
CHECK PUMP
STATUS
The PPC battery has been replaced. The PPC needs to check the
Pump status.
EMPTY RESERVOIR The PPC has recognized the Pump has 400 units (1 ml) or less
insulin remaining in its reservoir. Schedule a Pump refill as soon as
possible. Allowing the reservoir to completely empty may damage
the Pump.
HOURLY MAX
EXCEEDED
You attempt to deliver more than 2.5 times the bolus maximum in
one hour. To clear the message, press SEL and ACT. You may
exceed this limit by programming another bolus within 10 minutes.
LOW RESERVOIR The PPC has recognized the Pump has 800 units (2 ml) or less
insulin remaining in its reservoir. Schedule a Pump refill as soon as
possible. Allowing the reservoir to completely empty may damage
the Pump.
PPC LOW BATTERY The PPC can be programmed. Replace the PPC AA 1.5 volt
alkaline battery.
PUMP LOW
BATTERY
The Pump battery energy is low but still functioning. Schedule a
replacement as soon as possible.
PUMP RESET The Pump was reprogrammed to the preset factory values.
PUMP SELF TEST FAIL A Pump malfunction was detected during a Self Test. Notify your
physician immediately.
PUMP STOPPED
1 or 2 or 3 or 4 or 5 or 6
When the Pump recognizes a system malfunction, it automatically
stops and insulin delivery ceases. Notify your physician
immediately.
PUMP SUSPENDED The Pump is in suspend mode
TELEMETRY COMM
ERROR 3
The PPC and Pump are not communicating. Reposition the PPC
over the Pump, then press SEL and ACT. If the error message
persists, notify your physician immediately.