CHAPTER 8 PPC Care and Maintenance
Installing/replacing the main battery
PPC Battery life is estimated to be approximately 4 weeks. Only AA 1.5 volt
alkaline battery is recommended for use with the PPC. To replace the PPC
Main Battery:
1. Locate the battery door on the back of the PPC.
2. Slide the locking bar to the left.
3. Push the middle part of the PPC box (under the battery door) and lift by
gently pulling up the battery door to unlatch.
4. Remove the old battery, noting the polarity. The PPC screen will be blank.
5. Position the new battery so the + and - markings on the battery match the
polarity diagram in the battery compartment.
6. Close the battery door.
7. Slide the locking bar to the right.
8. The PPC screen reappears after 30 seconds.
9. The PPC will beep six times and display
“PPC/PUMP.”
10. After a few seconds, the screen will change to
“CHECK PUMP STATUS.”
11. Press SEL and ACT, then place the PPC over
the Pump.
PPC 8 102 100
PUMP 3 100 348
08:26 JAN O2
CHECK
PUMP STATUS