User's Manual

Table Of Contents
67
CHAPTER 9 Technical specifications
Implantable Insulin Pump (MMT-2007D)
Component Performance Specification
Diameter 8.1 cm (3.2 inches)
Thickness 2.0 cm (0.8 inches)
Reservoir Volume 13 ml to 15 ml
Weight - Empty 131 gm (4.6 ounces)
Insulin - Concentration Aventis HOE 21 PH U-400
Stroke Volume 0.42 to 0.58 µl per stroke
0.17 to 0.23 units per stroke
Basal Rate 0.2 to 35.0 units per hour (U/h)
Basal Patterns 3 Patterns of up to 48 basal rates each
Meal Bolus 0.2 to 25.0 units
Bolus Duration Immediate, Square Wave (30 minutes to 4 hours), or both together. Audio Bolus
Temporary Basal Rate 0.2 to 35.0 units per hour
30 minute increment duration
30 minutes up to 24 hours duration
Diagnostic Rate 10 to 150 U/h
Power Supply Lithium - Carbon Monofluoride Battery
Audio Alarms Low Battery
System Error
Safety Features Negative Pressure Reservoir with Passive Filling
Pump Shutdown and Alarm with System Error (unique code sequences)
Materials Titanium Housing, Polyethylene-lined Silicone Rubber Catheter