User Manual
Utilities 99 
Clear pump
The clear pump function resets the time and date and clears all pump settings except for language. The pump 
settings are restored to the factory defaults. When you clear your pump, you must reprogram all your settings 
before you can use your pump again. The pump does not clear the internal pump memory.   
Do these steps to clear your pump: 
1. Go to the CLEAR PUMP screen. Select 
Settings and press ACT.
Main > Utilities > Clear Pump
2. The CONFIRM screen will ask you to verify that you want to clear the 
pump. Select 
Yes then press ACT.
3. The RESET message will appear indicating the pump settings have been 
cleared. Read the instructions on the screen then press 
ESC, ACT to 
clear the message.
4. The pump will go through various screens while it restarts. 
After the pump clears all your settings, the screen will go to the 
TIME/DATE SETUP screen.
5. Reset the time and date as described in the section, “Setting the time and date” in chapter 3.
6. After you set the time and date, you must rewind your pump. Refer to the section, “Rewinding your pump” in 
chapter 4 for instructions. Remember, all your setting have been cleared and you must reprogram all your set-
tings.
WARNING: Do not clear your pump while it is connected to your body.
CAUTION: Do not clear your pump unless directed by your healthcare professional or a Medtronic MiniMed 
representative. If you clear your pump, it will be necessary to reprogram all your personal pump 
settings as directed by your healthcare professional. Additionally, you will have to rewind your 
pump.
MiniMed
CLEAR PUMP
Settings
ACT
MiniMed
CONFIRM
Yes
No
ACT
14:05
RESET
Cleared settings
Program as required
ESC, ACT to clear
MiniMed