Specications
Transmitter frequency 2.4 GHz
Guidance and manufacturers declaration
Table 1. Guidance and Manufacturers Declaration - Electromagnetic Emissions
This CareLink USB (MMT-7306) is intended for use in the electromagnetic environment specied below. The customer or the
user of the CareLink USB should make sure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions CISPR 11 Group 1
The CareLink USB uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not applicable
Table 2. Guidance and Manufacturers Declaration - Electromagnetic Immunity
This CareLink USB (MMT-7306) is intended for use in the electromagnetic environment specied below. The customer or the
user of the CareLink USB should make sure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
±4 kV contact ±4 kV contact
The CareLink USB should not be aected by electrostatic discharge that
might occur under normal conditions of use.
IEC 61000-4-2 ±8 kV air ±8 kV air
Table 2, continued next page