User Manual

System Maintenance 109
Table 6.4: Test Plug
Component Performance Specifications
Operating
Conditions
Temperature: 0 to +50 degrees Celsius
(+32 to +122 degrees Fahrenheit)
Relative Humidity: 15% to 95% with no
condensation
Storage
Conditions
Temperature: -20 to +55 degrees Celsius
(-4 to +131 degrees Fahrenheit)
Relative Humidity: 10% to 100% with
condensation
Approved
Cleaners
A mild household liquid detergent diluted with
water, a quaternary ammonia solution, or
isopropyl alcohol on a wipe.
(DO NOT place liquids into either connector.)
Approved
Disinfectants
Tap water, 10% bleach solution, 3% hydrogen
peroxide solution, 70% isopropyl alcohol.
Table 6.5: Transmitter
Component Performance Specifications
Biocompatibility Complies with ISO 10993-1 for long-term body contact
Operating
Conditions
Temperature: 0 to +50 degrees Celsius
(+32 to +122 degrees Fahrenheit)
Relative Humidity: 10% to 95% with no condensation
Transmitter Life 1 year under anticipated normal use conditions
Transmitter
Frequency
418 MHz (U.S. and Canada) or 433.92 MHz (International),
1mWerp, 20k5A1D Modulation