Getting Started 29
Setting the Glucose Unit
The Monitor gives you a choice of entering meter BG values and setting the low blood glucose and high
blood glucose limits in either mg/dl or mmol/l.
From the HOME screen:
1. Press the SEL button 7 times until the SETUP screen appears.
2. Press the ACT button once to go to the SETUP sub-screens.
3. Press the SEL button 5 times until the UNIT sub-screen appears.
4. Press the ACT button once, and use the Up or Down arrow button to choose either mg/dl or mmol/l.
5. Press the ACT button again to save the chosen units, and the Monitor will go to the ID sub-screen.
Press SEL 3 times to go to EXIT, and press ACT one to go back to the HOME screen.
GLUCOSE
UNIT
mg/dl
GLUCOSE
UNIT
mg/dl
SET