User Manual

72 Chapter 4
Low Batt
Error Message: LOW BATT (Low Batteries)
Cause: The voltage of the AAA batteries in the Monitor is low.
Action: Press the SEL button and then the ACT button to take care of the alarm. When this alarm is first
triggered, the batteries have about eight hours of normal operating life left. The time of the first alarm
is recorded into the Monitor memory. The Battery icon will be displayed on the HOME screen.
1. Turn Off the Monitor.
2. Change the batteries. See Chapter 2: Getting Started.
NOTE -
Make sure to put in new batteries within 10 minutes of taking out the old ones to avoid
automatic resetting of date and time. This will also avoid the triggering of a Replace Sensor alarm.
3. Do a manual Search procedure:
a. From the HOME Screen, press the SEL button 5 times until the SEARCH screen appears.
b. Press the ACT button and the Monitor will beep or vibrate. CONFIRM will be blinking on the
screen.
c. Press the ACT button again to begin a SEARCH. The screen will display SEARCH and the number 8
(minutes). A countdown will begin. It may take up to 8 minutes to finish the SEARCH.
NOTE -
Frequent use of the backlight and having the alert type on “vibrate” will use up batteries
quicker.
12:00
AM
LOW BATT
-
+