User's Manual

Table Of Contents
MySentry Monitor
Getting started 15
Getting started 3
Setting up communication
Before you can use the MySentry system, you must set up communication from the pump to the
Monitor, and the Monitor to the Outpost (if used). The following linking procedures only need to be
performed once, unless the Monitor or Outpost is unplugged or an alarm has cleared the pump
settings. If so, then you must re-link the Monitor to the pump.
NOTE: It is recommended that you place the pump, Monitor, and Outpost in one location during
the setup process, and you not move any component until they are linked. After the components
are successfully linked, you can move them to their permanent locations. The Monitor and
Outpost work best if they stay in one location in your home.
Linking the Monitor to the pump
Do not move the pump or Monitor during this process.
1 Remove the pump from the user and place it within six feet (183 cm) of the Monitor.
2 Use the power supply provided with the MySentry system to plug the Monitor into a standard electrical
wall outlet.
3 If the Monitor is not linked to a pump, the Link to Pump screen displays.
If the Link to Pump screen does not display, follow these steps:
a. Press until the main menu displays.
b. Go to the Link to Pump screen.
Main Menu > Go to Utilities Menu > Link to Pump