User's Guide
12
Guidance and manufacturer’s Declaration-Electromagnetic Immunity
Immunity Test IEC 60601-1-
2:2014
Test Level
IEC 60601-1-
2:2014
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
discharge (ESD)
±8kV contact
±2, ±4, ±8, ±15
kV air
±8kV contact
±2, ±4, ±8,
±15kV air
For use in a
typical domestic,
commercial,
or hospital
environment.
Conducted
disturbances
induced by RF
fields
3Vrms
150kHz to 80MHz
6Vrms
ISM bands
between 150 kHz
to 80MHz
Not
applicable
Requirement does
not apply to this
battery powered
device.
Electrical fast
transient/burst
±2kV
100kHz repetition
frequency
Not
applicable
Requirement does
not apply to this
battery powered
device.
Surge Line to Line: ±0.5
kV,±1kV
Line to Ground:
±0.5kV, ±1kV,
±2kV
Not
applicable
Requirement does
not apply to this
battery powered
device.
Note: U
T
is the a.c. mains voltage prior to application of the test level.