Guidance and manufacturer’s Declaration-Electromagnetic Immunity
Immunity Test IEC 60601-1-
2:2014
Test Level
IEC 60601-1-
2:2014
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
discharge (ESD)
±8kV contact
±2, ±4, ±8, ±15
kV air
±8kV contact
±2, ±4, ±8,
±15kV air
For use in a
typical domestic,
commercial,
or hospital
environment.
Conducted
disturbances
induced by RF
fields
3Vrms
150kHz to 80MHz
6Vrms
ISM bands
between 150 kHz
to 80MHz
Not
applicable
Requirement does
not apply to this
battery powered
device.
Electrical fast
transient/burst
±2kV
100kHz repetition
frequency
Not
applicable
Requirement does
not apply to this
battery powered
device.
Surge Line to Line: ±0.5
kV,±1kV
Line to Ground:
±0.5kV, ±1kV,
±2kV
Not
applicable
Requirement does
not apply to this
battery powered
device.