Guidance and manufacturer’s Declaration-Electromagnetic Immunity
Radiated RF 10V/m
80MHz to 2.7GHz
80%AM at 1kHz
10V/m
80MHz to
6GHz
80%AM at
1kHz
Portable and mobile
RF communications
equipment should
be used no closer
to any part of the
transmitter than
the recommended
separation distance
of 30cm (12 in).
Field strengths from
fixed RF transmitters,
as determined by
an electromagnetic
site survey, should
be less than the
compliance level
in each frequency
range.
Interference may
occur in the vicinity
of equipment
marked with the
following symbol:
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption, and reflection from structures,
objects and people.
Maintenance
Operation
Operating temperature range 36°F to 104°F (2°C to 40°C)
Air pressure range 700 hPa to 1060 hPa (10.2 psi to 15.4 psi)
Operating relative humidity (RH) range 15% to 95%