User's Guide
7
Inserting the Simplera™ sensor
YYYY MM DD
XXXX XXX XXX
CODE: XXXXXX
1
The inserter label is on the top of the inserter.
1. Before insertion, perform the following steps:
a. Inspect the expiration date. Do not use an expired
Simplera™ sensor.
b. Make note of the serial number (SN) and the CODE.
Both numbers will be used later to pair the sensor
with the Simplera™ app.
Note: The SN and CODE label is also on the inside of the
Simplera™ sensor box lid.
2
2. Inspect the cap label for damage before insertion.
Note: Do not use the Simplera™ sensor if the cap label is
damaged or missing from the cap.
3
3. Inspect the tamper band to make sure that it is not
broken, damaged, or missing from the device.
Note: Do not use the Simplera™ sensor if the tamper
band is broken, damaged, or missing from the
device.
4
4. Wash hands thoroughly with soap and water.
Note: Wear gloves when inserting the Simplera™ sensor
into another person to avoid accidental contact
with patient blood. Minimal bleeding can occur.