How-To Guide

Auto Mode exit alert
•High SG alert
For more information about the Auto Mode exit alert or the High SG alert,
see SmartGuard Auto Mode alerts and messages, on page 273. For more
information about the Low SG XX mg/dL (XX represents 50 mg/dL or
below) alarm, see CGM (sensor) alarms, alerts, and messages, on page 262.
You can check the status of the Alert Silence option in the Sensor screen. For more
information, see Status screens, on page 50.
The following table describes the glucose alerts that are silenced with each option.
Option Silences these alerts
High Alerts
Only
Alert on high, Alert before high, and Rise Alert
High and
Low Alerts
Alert on high, Alert before high, Rise Alert, Alert on low, and Alert
before low
Note: Alert on low cannot be silenced if the
SmartGuard Suspend on low or SmartGuard Suspend
before low options are turned on.
All Sensor
Alerts
All of the alerts listed previously for High and Low Alerts, plus the
following:
All calibration alerts, reminders, or error messages
All alerts relating to sensor insertion, including alerts about sensor
warm-up, changing your sensor, sensor expiration, sensor errors,
connection issues, and so on
All alerts related to your transmitter, including all alerts about your
transmitter battery and all connection issues
To set Alert Silence in Auto Mode:
1. Press
and go to the Alert Silence screen.
Audio Options > Alert Silence Options
SmartGuard Auto Mode
SmartGuard Auto Mode 235