Title and text Type Explanation Next steps
Device Limit
You must
delete an
existing device
(device type)
before you can
pair a new one
(device type).
Message The pump is already paired
with the maximum number
of devices for this type.
The following list describes
the maximum number of
each device type to pair
with the pump:
• Meter–four Accu-Chek
Guide Link meters
• CGM–one Guardian Link
(3) transmitter
• Mobile Device–one
compatible mobile device
• Select OK to clear the
message.
• Go to the Manage
Devices screen and
select the device you
want to delete from the
list of devices.
Select Delete, and then
select Yes to confirm or
No to cancel.
Pair the pump and the
desired device.
Device not
compatible
Device cannot
be used with
this pump.
Alert The pump cannot pair with
the selected device.
• Select OK to clear the
alert.
• Call 24-Hour Technical
Support for assistance.
248 Chapter 13