Title and text Type Explanation Next steps
Device not
found
Make sure
device is in
range and in
pairing mode.
Alert The pump did not pair
with the device.
• Select OK to clear the
alert.
• Make sure the device is
not already paired with
a pump.
• Make sure the device is
ready to pair with the
pump.
• Make sure you are away
from any electronic
devices that might cause
interference, such as
cellular phones that are
not paired with the
MiniMed 770G System
and other wireless
devices.
• Move the device closer
to the pump.
• Try to pair the pump
with the device again.
Fill Cannula?
Select Fill to fill
cannula or
select Done if
not needed.
Alarm You had the Fill Cannula
screen displayed for
15 minutes.
• To proceed and fill the
cannula, select Fill.
• If you do not need to fill
the cannula, select Done
to skip this process.
■
Alarms, alerts, and messages
Alarms, alerts, and messages 249