How-To Guide

Title and text Type Explanation Next steps
Low battery
Pump
Replace battery
soon.
Alert The battery in the pump is
low on power. Remaining
battery life is 10 hours or
less.
Select OK to clear the
alert.
Replace the AA battery
as soon as possible.
Otherwise, insulin
delivery stops, and the
Replace Battery Now
alarm occurs.
•If the pump is delivering
a bolus or filling the
cannula, wait until
delivery is complete to
replace battery.
Low BG XX
mg/dL
Treat Low BG.
Do not bolus
until BG is
normal. Monitor
BG. Confirm
BG?
Alert Your BG meter reading is
below 70 mg/dL.
Select No to prevent the
remote BG from being
used by your pump.
Select Yes to confirm
the BG reading.
Check your BG and treat
as necessary.
Low reservoir
XX units
remaining.
Change
reservoir.
Alert Your reservoir is low on
insulin, according to the
number of units set in the
Low Reservoir Reminder.
Select OK to clear the
alert.
Change the reservoir
soon.
If you do not change
the reservoir after you
receive this alert, you
will receive a second
Low reservoir alert when
the insulin level reaches
half of your original alert
amount.