Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
% Insulin Remaining
Time (hours)
Active Insulin Curve
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
01 23456 978
8 Hour
7 Hour
6 Hour
5 Hour
4 Hour
3 Hour
2 Hour
Graph adapted from Mudaliar and colleagues, Diabetes Care, Volume 22,
Number 9, Sept. 1999, page 1501.
Carb ratios
Maximum ratio settings Range
8 1 to 200 grams/unit
Delivery accuracy
• For a basal rate of 1.0 U/h, the delivery accuracy is ±5%.
For a basal rate of 0.025 U/h, the delivery accuracy is ±10%.
Delivery accuracy for bolus volumes < 0.1 unit is ±20% and delivery accuracy
for bolus volumes ≥ 0.1 unit is ±5%.
• All Normal boluses are delivered within 16 minutes, 41 seconds ±3 seconds at
Standard rate (25 units, at 1.5 units per minute), and within 1 minute,
41 seconds ±3 seconds at Quick rate (25 units, at 15 units per minute).