Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MiniMed 770G insulin pump is intended for use in the electromagnetic
environment specified below. The customer or the user of the MiniMed 770G insulin
pump should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment -
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2,
60601-1-2
±8 kV contact
±2, 4, 8, 15 kV
air
±8 kV contact
±2, 4, 8, 15 kV
air
For use in a typical
domestic, commercial,
or hospital
environment.
Conducted
disturbances induced
by RF fields
3 Vrms
150 kHz to
80 MHz
6 Vrms
ISM bands
between
150 kHz to
80 MHz
Not applicable Requirement does not
apply to this battery
powered device.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV
100 kHz
repetition
frequency
Not applicable Requirement does not
apply to this battery
powered device.
Surge
IEC 61000-4-5
Line to Line:
±0.5 kV, ±1 kV
Line to Ground:
±0.5 kV, ±1 kV,
±2 kV
Not applicable Requirement does not
apply to this battery
powered device.
318 Chapter 16