Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MiniMed 770G insulin pump is intended for use in the electromagnetic
environment specified below. The customer or user of the MiniMed 770G insulin
pump should assure that it is used in such an electromagnetic environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Radiated RF
IEC
61000-4-3
IEC
60601-1-2
10 V/m
80 MHz to
2.7 GHz
80% AM at 1
kHz
10 V/m
80 MHz to
2.7 GHz
80% AM at 1
kHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
MiniMed 770G insulin pump, including
cables, than the recommended
separation distance of 12 inches (30
cm).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note: The table is per IEC (EN) 60601-1-2 Edition 4.
Icon glossary
For a definition of the symbols displayed on the device and package labels, please
see www.medtronicdiabetes.com/symbol-definitions.
320 Chapter 16