Product Info

Table Of Contents
Table D-1: Run-In Period Severe Hyperglycemia:
Cause Total
Infusion set change 9
Occlusion Alarm 3
Infusion set fell out 2
Bent or Kinked Cannula 1
Total 15
Table D-2: Study Period Severe Hyperglycemia:
Cause Total
Infusion set change 28
Occlusion Alarm 12
Infusion set fell out 7
Bent or Kinked Cannula 5
Infusion set change or safe basal 3
Safe basal 2
Suspend before low suspension 1
Automatic & manual suspensions 1
Unclipped infusion set 1
Internal Battery Connector Resistance 1
Manual suspension and safe basal 1
Total 62
The following table shows the time spent per day in specific glucose ranges
during the run-in and study phases by all subjects.
Table D-3: Time Spent in Specific Glucose Ranges During the Run-In and Study Phases by All Subjects
Glucose Range (mg/dL) Run-In Phase Study Phase
Time in Glucose Range (min) Mean±SD Time in Glucose Range (min) Mean±SD
≤50 12.2±16.9 7.8±7.3
≤60 32.3±32.7 20.3±14.2
≤70 68.3±55.4 43.1±23.6
70–180 808.6±163.5 936.2±110.4
>180 563.1±184.3 460.7±110.5
>250 191.0±111.5 148.4±74.1
>300 68.1±54.8 53.7±39.4
>350 23.0±24.9 17.4±17.1
The following table shows the ranges of changes in HbA1C observed in the study
and indicates the number of subjects that demonstrated each type of change in
HbA1C observed.
670G Performance Data and Technical Information
670G Performance Data and Technical Information 351