Product Info
Table Of Contents
- M998590A011DOC_A_FINAL_UserGuide_Part2
- Troubleshooting
- Maintenance
- Product specifications and safety information
- Product specifications
- Alarm and alert escalation
- Altitude range
- Audio frequency
- Backlight
- Basal delivery
- BG Target
- BG meter value
- Bolus delivery
- Bolus Wizard feature default settings
- Bolus Wizard feature specifications
- Carb ratios
- Delivery accuracy
- Easy Bolus feature
- Environmental conditions
- Essential performance
- Filling the infusion set and cannula
- Infusion pressure
- Insulin delivery default settings
- Bolus settings
- Basal settings
- Insulin sensitivity factor
- Low Reservoir reminder
- Max Bolus
- Normal bolus
- Occlusion detection
- Percent temp basal
- Program safety checks
- Pump dimensions
- Pump memory
- Pump weight
- Sensor default settings
- Wireless communication
- FCC notice
- IEC60601-1-2:4th Edition notice
- Guidance and manufacturer's declaration
- Icon glossary
- Product specifications
- 670G Performance Data and Technical Information
Table D-1: Run-In Period Severe Hyperglycemia:
Cause Total
Infusion set change 9
Occlusion Alarm 3
Infusion set fell out 2
Bent or Kinked Cannula 1
Total 15
Table D-2: Study Period Severe Hyperglycemia:
Cause Total
Infusion set change 28
Occlusion Alarm 12
Infusion set fell out 7
Bent or Kinked Cannula 5
Infusion set change or safe basal 3
Safe basal 2
Suspend before low suspension 1
Automatic & manual suspensions 1
Unclipped infusion set 1
Internal Battery Connector Resistance 1
Manual suspension and safe basal 1
Total 62
The following table shows the time spent per day in specific glucose ranges
during the run-in and study phases by all subjects.
Table D-3: Time Spent in Specific Glucose Ranges During the Run-In and Study Phases by All Subjects
Glucose Range (mg/dL) Run-In Phase Study Phase
Time in Glucose Range (min) Mean±SD Time in Glucose Range (min) Mean±SD
≤50 12.2±16.9 7.8±7.3
≤60 32.3±32.7 20.3±14.2
≤70 68.3±55.4 43.1±23.6
70–180 808.6±163.5 936.2±110.4
>180 563.1±184.3 460.7±110.5
>250 191.0±111.5 148.4±74.1
>300 68.1±54.8 53.7±39.4
>350 23.0±24.9 17.4±17.1
The following table shows the ranges of changes in HbA1C observed in the study
and indicates the number of subjects that demonstrated each type of change in
HbA1C observed.
■
670G Performance Data and Technical Information
670G Performance Data and Technical Information 351