Users Manual

WARNING: Always check your tubing for air bubbles. Continue to
press Fill until the bubbles have been removed from the tubing. Air
bubbles may result in inaccurate insulin delivery.
To fill the tubing:
1. After you load your reservoir and select Next from the Load Reservoir screen,
the Fill Tubing screen appears.
2. Select and hold Fill. Your pump beeps six times as it dispenses insulin into
the tubing toward the infusion set needle. Continue to hold Fill until insulin
droplets form on the tip of the infusion set needle, and then release. Your
pump beeps as it fills the tubing, and the amount of insulin used appears on
the screen.
If the Max Fill reached alarm occurs, it means you have used more than
30 units of insulin to fill your tubing. For details, go to Pump alarms, alerts, and
messages, on page 244, and see the description for Max Fill reached.
3. Select Next to continue.
4. Follow the instructions in the next section to insert the infusion set into your
body before filling the cannula.
Inserting the infusion set
WARNING: Do not remove the reservoir from the pump while the
infusion set is connected to your body. Doing so could result in the
delivery of too little or too much insulin, which can cause high BG
or low BG.
126 Chapter 5