All connections to the Biopac MP150 interface are made using cables provided
by Medtronic.
3.3 CONNECTIONS TO AND USE OF THE AHM MODULE
Table 3 depicts channel connections to the AHM Module’s rear (Figure 4), and
front panels (Figure 5). The LV TIP, and LV RING 1-3 connections only apply
when an LV catheter is placed. When an LV catheter is placed, select the distal
electrode on the LV catheter as electrode AA. Select electrodes BB, CC, and DD
such that the edge-to-edge distance between BB and AA is closest to 21mm,
between CC and BB is closest to 1.3mm, and between DD and AA is closest to
46mm.
Table 3. Connections
to the AHM Module
Connection Method
HVB/LVR3 Cable to Biopac UIM100C Analog Input Channel 5
I BLANK Cable to Biopac UIM100C Analog Input Channel 6
AOUT Cable to Biopac UIM100C Analog Input Channel 7
GAIN Cable to Biopac UIM100C Analog Input Channel 8
PATIENT CABLE Model 15420 Patient Cable to sterile field for connection to subject
pacing leads
LV TIP Cable to LV catheter electrode AA (if LV catheter is placed)
LV RING 1 Cable to LV catheter electrode BB (if LV catheter is placed)
LV RING 2 Cable to LV catheter electrode CC (if LV catheter is placed)
LV RING 3 Cable to LV catheter electrode DD (if LV catheter is placed)
Model 2067 or 2067L
Programming Head
Programmer Head for Model 2090 Programmer
Except for the LV catheter, all connections to the AHM Module are made using
cables provided by Medtronic. LV catheter cables are provided by the LV
catheter manufacturer.