8.1 Cellular connection error
• Error: The cellular signal icon lights up orange,
and the bottom segment of the light ring flashes
orange. The communicator plays an error tone
(1 high note and 1 low note play twice).
• Cause: The communicator does not have an
adequate cellular connection.
• Action: Unplug the communicator and move it
to a place where it might receive a better cellular
signal (for example, next to a window).
If that does not resolve the error, connect the
communicator to WiFi using the instructions in
Chapter 5, “Optional: Connecting the
communicator to WiFi”, page 10.