User's Manual

Table 2. Patient Connector specifications (continued)
Electrical safety requirements per IEC 60601-1:2005, Clause 8.7
Touch leakage current ≤0.1 mA
Patient leakage current A.C. ≤0.1 mA
Patient auxiliary current
D.C.
≤ 0.01 mA
Patient auxiliary current
A.C.
≤0.1 mA
IEC 60529 Degrees of Protection Provided by Enclosures (IP Code) - AMD
7643: July 1993; AMD 10931: August 2000
Ingress This product complies with international electrical
safety rating IP22 with regard to ingress of dust,
other foreign objects, and water as required by IEC
60601-1.
Physical dimension and weight
Height
Width
Depth
Weight
14 cm
7.2 cm
3.2 cm
.18 kg
Temperature limits
Operating
Storage
Transport
15 °C to 35 °C
- 20 °C to 60 °C
- 30 °C to 70 °C
Humidity limits
Operating
Storage
Storage Transport
75%
95% at 35 °C
93% at 35 °C
Altitude
Maximum 3000 m
Connectivity and controls
Bluetooth 2.1 and 4.0
Frequency range 2.4 GHz
Output power Less than 10 mW EIRP
Expected service life
5 years
4.3.1 Expected service life
The patient connector has an expected service life of 5 years. If you notice
decreased operational time between battery charges or the battery no longer holds
its charge, contact Medtronic to get a replacement patient connector.
4.3.2 Disposal of the patient connector
Return the patient connector to Medtronic for proper disposal.
16
M999999A Medtronic Confidential DRAFT Composed: 2014-11-20 11:53:03
XSL-Stylesheet J - Size-selectable package
manual 24-JAN-2014