7. Wrap the blue weight around the patient connector and insert the screw into the screw hole. Use the Medtronic-supplied wrench
to tighten it.
4.3 Software updates
Software updates will be pushed automatically to the patient connector when you start a communication session in the app. Updates
can take up to 5 minutes. You must wait until the update is complete before you can continue with the session.
4.4 Specifications
Table 2. Patient connector specifications
Standards (The patient connector complies with the following:)
Radio frequency wireless specifications and applicable standards
EMC EN / IEC 60601-1-2
EN 300 328
EN 301 489
EN 302 195
EN 55011 Class B
EN 55024 Class B
Radio RSP-100
RSS-GEN
RSS-243
RSS-310
RED 2014/53/EU
Patient safety UL/CUL 60601-1, Type BF applied part
EN 60601-1, Class 2, continuous operation, Type BF
AC power requirement
Voltage 100–240 VAC nominal
Frequency 50/60 Hz nominal
Battery
Type Li-polymer, rechargeable
Capacity 1500 mAh
Charge duration Standby: 15 days
Operating: 2 hours (typical)
Voltage 3.7 V
Patient connector charger specifications
Power supply
Model MENB1020A0500XXX power supply (Medtronic re-order number 249651)
Voltage in 100-240 VAC 0.5A at 50-60 Hz
Voltage out 5 VDC 3 A
Electrical shock protection class Class II
Intended duty Continuous
USB cable
Model 249702 USB cable
Voltage 5 V 0.8 A
Power 4W
24970A charge cradle
Model 24970A
Voltage 5 V 3.0 A
Printing instructions: doc #163256
Category: Size-selectable pkg
Size: 8.5 x 11 in (216 x 279 mm)
502386-031
23
M962083A001 C Medtronic Confidential Composed: 2016-08-16 12:05:16