Operating instructions

LIFEPAK CR Plus Defibrillator Operating Instructions 1-5
©2002–2005 Medtronic Emergency Response Systems, Inc.
1 Introduction
Introduction
Note: Medtronic Emergency Response Systems devices, electrodes, and cables are latex-free.
Possible electrical interference with device performance.
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference
(RFI) which could affect the performance of this device. RFI may result in improper device operation,
distorted ECG, failure to detect a shockable rhythm, or cessation of pacing. Avoid operating the device
near cauterizers, diathermy equipment, cellular phones, or other portable and mobile RF
communications equipment. Maintain equipment separation of at least 1.2 m (4 ft) and do not rapidly
key EMS radios on and off. Contact a technical support representative if assistance is required.
Possible electrical interference.
Using cables, electrodes, or accessories not specified for use with this device may result in increased
emissions or decreased resistance to electromagnetic interference which could affect the performance
of this device or of equipment in close proximity. Use only parts and accessories specified in these
operating instructions.
Possible electrical interference.
This defibrillator may cause electromagnetic interference (EMI) especially during charge and energy
transfers. EMI may affect the performance of equipment operating in close proximity. If possible, verify
the effects of defibrillator discharge on other equipment prior to using the defibrillator in an emergency
situation.
Possible improper device performance.
Using other manufacturers’ cables or electrodes may cause the defibrillator to perform improperly and
invalidates the safety agency certification. Use only the parts and accessories specified in these
Operating Instructions.
Safety risk and possible equipment damage.
Monitors, defibrillators, and their accessories (including electrodes and cables) contain ferromagnetic
materials. As with all ferromagnetic equipment, these products must not be used in the presence of the
high magnetic field created by a Magnetic Resonance Imaging (MRI) device. The high magnetic field
created by an MRI device will attract the equipment with a force sufficient to cause death or serious
personal injury to persons between the equipment and the MRI device. This magnetic attraction may
also damage the equipment. Skin burns will also occur due to heating of electrically conductive
materials, such as patient leads and pulse oximeter sensors. Consult the MRI manufacturer for more
information.
Shock hazard.
Do not insert a finger or any object other than the CHARGE-PAK
into the well of the defibrillator.
CAUTION!
Possible equipment damage.
This defibrillator may be damaged by mechanical or physical abuse, such as immersion in water or
dropping the defibrillator. If the defibrillator has been abused, remove it from use and contact a qualified
service technician.
WARNINGS!