Users Manual

163
Note: The Connect Sensor option is only available when no Sensor is currently
connected to the PDM.
2. Make sure that your Transmitter is connected to a Sensor and that your
Transmitter SN is found or entered, and then continue by tapping Next.
3. If the sensor is calibration-free, enter the sensor code on the sensor label
which is unique for each sensor. Once the sensor code is entered
successfully, the calibrations aren’t required.
Or, skip the sensor code input step and go to Connection screen. The
sensor need to be calibrated twice (once every 12 hours) on the first day
and then once every 24 hours since the second day.
4. When finished, the following screen appears.