Users Manual

165
Note: If your sensor is not a calibration-free sensor, or you skip the sensor
code input step, you must calibrate your sensor at least twice (once every 12
hours) on the first day and then once every 24 hours since the second day. If
you have entered sensor code successfully, the system won’t require
calibration. But you can calibrate the sensor if you want.
Note: Calibration is unavailable under the following circumstances:
Sensor disconnected from the PDM
Sensor warm-up
Within 15 min after the alert SENSOR CAL ERROR
Poor RF communication between the Transmitter and the PDM
No Readings
6.3.1 Enter Your Meter BG
Here you can enter your present blood glucose measured by a finger prick
blood glucose meter.
1. Go to the Enter BG screen.
Main MenuSensorSensor Calibration