167
6.3.2 Set Calibration Repeat
Go to the Cal Repeat screen.
Main MenuSettingsCGM SystemCal Repeat
After you receive and clear a “METER BG NOW” alert, PDM will repeat the
alert until you enter a new blood glucose measurement.
You can turn Cal Repeat on/off. If Cal Repeat is on, you can set the repeat time
of “METER BG NOW” alert from 5 min to 1h with an increment of 5 min.
6.3.3 Calibration Reminder
Calibration reminder enables you to get reminded a certain time before the
due time of next calibration.
1. Go to the Cal Reminder screen.