Interacting with your PDM includes activities such as pressing the power
button and checking the screen.
With the Alert Silence feature you can keep glucose alerts silent for a specified
time of 30 minutes to 24 hours.
There are five Alert Silence options:
Off This means all glucose alerts are turned on: the PDM will beep or
vibrate if any Sensor alert occurs.
Low The PDM will not beep or vibrate if a low alert (LOW GLUCOSE,
RAPID FALL or LOW PREDICTED) occurs during the specified time.
High The PDM will not beep or vibrate if a high alert (HIGH GLUCOSE,
RAPID RISE or HIGH PREDICTED) occurs during the specified time.
High and Low The PDM will not beep or vibrate if a high/low alert
(HIGH/LOW GLUCOSE, RAPID RISE/FALL, HIGH/LOW PREDICTED) occurs
during the specified time.
All The PDM will not beep or vibrate if “LOST SENSOR”, “SENSOR CAL
REMINDER”, METER BG NOW”, “SENSOR EXP IN 6 HOURS”, “SENSOR EXP
IN 2 HOURS”, “SENSOR EXP IN 30 MINS”, “SENSOR EXPIRED”, or any of
the high/low alert occurs during the specified time.
See “Status Bar Icons” and Chapter “Safety System and Alarms/Alerts” for
more information.
6.4.6 Sensor Expired
Go to the Sensor Expired screen.
Main MenuSettingsCGM SystemSensor Expired