Average SG: Average SG readings of the selected days.
Time in target range: The percentage of the duration in which SG reading is in
the target range (3.9 - 10.0 mmol/L or 70 - 180 mg/dL).
Time above range: The percentage of the duration in which SG reading is above
the target range (10.0 mmol/L or 180 mg/dL).
Time below range: The percentage of the duration in which SG reading is below
the target range (3.9 mmol/L or 70 mg/dL).
6.6 Troubleshooting CGM issues
Can I take a sauna with the CGM System on?
No.
Firstly, the operating temperature range for the Transmitter is +5°C ~ +40°C.
Secondly, if you take a sauna, your blood glucose can fluctuate.
Can I dive with a Sensor on?
No.
Your Sensor (including the installed Transmitter) is waterproof to a depth of 2.5
meters (8 feet) for up to 60 minutes (IPX8).
It means the maximum pressure the device can tolerate equals the pressure in
2.5m deep in STILL water instead of flowing water.
It is OK to take a shower or go swimming with the devices on, but if you go diving,
the water pressure may be too high for the devices.