Users Manual

177
I didn’t see an alert message, but it appeared in History.
If one of the following alerts happened, the PDM would beep/vibrate and
display a message first, and if you missed that alert, later when you checked the
PDM, the condition that triggered the alert had changed (for example, your
glucose level returned to the target range), then you wouldn't see any message
on the screen, you would only find it in History.
1. LOW GLUCOSE
2. HIGH GLUCOSE
3. LOW PREDICTED
4. HIGH PREDICTED
5. RAPID RISE
6. RAPID FALL
7. ALERT SILENCE
8. SENSOR ERROR
9. BELOW 3.1 mmol/L (56 mg/dL)
10. LOST SENSOR
If the following alert happened, the PDM would beep/vibrate and display a
message first, and if you missed that alert, later when you checked the PDM,
the alert may have escalated to another alert/alarm, and you will ONLY see the
message of the escalated alert/alarm. The first alert will appear in History.
Alert Alert Escalation
SENSOR EXP IN 6 HOURS
SENSOR EXP IN 2 HOURS, then SENSOR
EXP IN 30 MIN, at last SENSOR EXPIRED
Charging the Transmitter
We recommend that you charge the Transmitter after each Sensor session, or
make sure that at least 1 minute has passed before you attach the Transmitter
to a new Sensor.
Green lights after installing the Transmitter
After you install the Transmitter, the green light on the Transmitter will flash 3
times immediately indicating that the Transmitter is properly connected with
the Sensor, and flash another 6 times within one minute indicating that the
system check has completed.
B ureau Veritas