Users Manual

193
Main MenuEasyLoop(Pre) Low Suspend
Note: You can set the Time before Low between 5 min and 40 min with an
increment of 5 min. The factory default is 30 min.
Triggering Conditions for Predictive Low Suspend (from 30 min to 2 h after
suspension)
Both of the following two conditions must be met to start Predictive Low
Suspend.
The Sensor glucose value is at or within 3.9 mmol/L (70 mg/dL) above the
low suspend limit.
The Sensor glucose value is predicted to fall at or within 0.8 mmol/L (15
mg/dL) above the low suspend limit in the set period of time and the rate
of glucose change is negative.
Time of Suspension
Once Predictive Low Suspend is triggered, the period of suspension will last for
at least 30 minutes unless you manually resume basal insulin. The maximum
suspension time is 2 hours. After 2 hours of suspension, basal insulin will be
resumed unconditionally.
Triggering Conditions for Automatic Resumption of Basal
Both of the following two conditions must be met for the system to resume
basal insulin automatically.
The Sensor glucose value is at least 0.8 mmol/L (15 mg/dL) higher than the
low suspend limit.