The Sensor glucose value is predicted to be at least 1.7 mmol/L(30 mg/dL)
higher than the low suspend limit in half an hour.
Reminder of Resumption
Whether a Predictive Low Suspend alert is cleared or not, the same Reminder
will appear when insulin is automatically resumed.
When the Low Suspend Feature and Predictive Low Suspend feature are
Unavailable
After insulin delivery is resumed from Low Suspend or Predictive Low Suspend,
the Low Suspend feature and Predictive Low Suspend feature will be
unavailable for 30 minutes.
7.2 Summary History: Low Suspend History
Go to Low Suspend History screen.
Main MenuHistorySummary HistoryLow Suspend History