Alarm sound wave:
Icon Sound wave Significance
PDM emits ten
beeps/vibrations each time.
PDM emits Three
beeps/vibrations each time.
Patch Pump emits Three
beeps/vibrations each time.
Patch Pump emits Three
beeps/vibrations each time.
8.3.1 PDM Alarms
If a PDM alarm is not cleared within 10 minutes, your PDM will make a siren
sound until the alarm is cleared.
PDM Message Priority Reason Actions to Take
PDM ERROR
Remove device.
Call customer
support.
A PDM error is
detected.
Tap to clear it.
Remove Pump and
Sensor.
Contact customer
support immediately.
Check blood glucose.
PDM ERROR
The PDM has
restarted. Change
patch.
A PDM software
error is detected
and the PDM has
restarted, but no
settings have
been changed.
Tap to clear it.
Remove the Patch
Pump and change a
Reservoir Patch.
If the problem occurs
repeatedly, please
contact customer
support.
CHARGE PDM
NOW
Charge PDM now.
The PDM battery
is depleted.
Tap to clear it.
Charge PDM battery.