8.3.2 Pump Alarms
When a Pump alarm occurs:
Indicator light: The indicator light on the pump flashes red once per second until
the alarm is cleared.
Note: If a Pump alarm is not cleared within 10 minutes, both your PDM and
Patch Pump will make a siren sound until the alarm is cleared.
The following table lists high priority alarm messages.
PDM Message Priority Reason Actions to Take
OCCLUSION
DETECTED
Delivery
stopped. Change
Patch now.
Pump occlusion
is detected.
Tap to clear it.
Change Patch.
Check blood glucose.
PATCH ERROR
Delivery
stopped. Change
Patch now.
A Reservoir
Patch error is
detected.
Tap to clear it.
Change Patch.
Check blood glucose.
PUMP BASE
ERROR
Remove Pump.
Call customer
support.
A Pump Base
error is detected.
Tap to clear it.
Remove Pump.
Contact customer
support immediately.
Check blood glucose.
The following table lists medium priority alarm messages.
PDM Message Priority Reason Actions to Take
AUTO OFF
Delivery
suspended. No
status received.
The PDM has not
received a pump
status during the
time limit set.
Tap to clear it.
Resume basal delivery.
Check blood glucose
and treat it as
necessary.
Check Pump history.