PDM Message Priority Reason Actions to Take
PATCH EXPIRED
Delivery will
stop. Change
Patch now.
The current
Reservoir Patch
has reached the
end of its 3-day
operating life.
Tap to clear it.
Change Patch.
Check blood glucose.
PATCH BATT
DEPLETED
Delivery
stopped. Change
Patch now.
The Reservoir
Patch battery is
depleted.
Tap to clear it.
Change Patch.
Check blood glucose.
EXCEEDS MAX
TDD
Exceeds max
TDD. Delivery
stopped.
You have
attempted to
deliver more
insulin than
expected based
on your Daily
Max setting.
Tap to clear it.
Check blood glucose.
Resume basal delivery.
Check bolus history and
reevaluate your need
for insulin.
Continue to monitor
blood glucose.
EXCEEDS MAX
1HR DELIVERY
Exceeds 1 hour
max. Delivery
stopped.
You have
attempted to
deliver more
insulin than
expected based
on your Hour
Max setting.
Tap to clear it.
Check blood glucose.
Resume basal delivery.
Check bolus history and
reevaluate your need
for insulin.
Continue to monitor
blood glucose.
EMPTY
RESERVOIR
Delivery
stopped. Change
Patch now.
There is no
insulin in the
reservoir.
Tap to clear it.
Change Patch.
Check blood glucose.
LOW SUSPEND
Low glucose
suspend
activated.
The last Sensor
glucose reading
is at or below the
Low Glucose
Suspend Limit
set.
Tap to clear it.
Check blood glucose
and treat it as
necessary.