Users Manual

215
9 Manufacturer’s Declaration
9.1 Electromagnetic Emissions
Emissions Test Compliance
RF emissions
EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR
11:2009+A1:2010 and IEC 60601-1-2:2014
Group 1
RF emissions
EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR
11:2009+A1:2010 and IEC 60601-1-2:2014
Class B
9.2 Electromagnetic Immunity
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment
The A7+ TouchCare® System is intended for use in the electromagnetic
environment specified below. The customer or user of the A7+
TouchCare® System should assure that it is used in such an
electromagnetic environment.
Electrostatic
Discharge IEC
61000-4-2
±2kV, ±4kV,
±8kV contact
discharge
±2kV,±4kV,±8k
V, ±15kV air
discharge
±2kV, ±4kV,
±8kV
contact
discharge
±2kV,±4kV,
±8kV,
±15kV air
discharge
For home healthcare
environment and
professional healthcare
facility environment
Power Port,
Signal and
Iterconnecting
Cable
Table 5 of IEC
60601-1:2014
±2Kv
100Hz,
During the
test, the
EUT can
The network power
supply should have the
quality used in a typical
B ureau Veritas