Users Manual

216
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment
The A7+ TouchCare® System is intended for use in the electromagnetic
environment specified below. The customer or user of the A7+
TouchCare® System should assure that it is used in such an
electromagnetic environment.
IEC 61000-4-
4:2012
for AC power
port
operate as
intended
commercial or hospital
environment.
Surge
IEC 61000-4-
5:2005
±0.5kV, ±1kV
(different
mode)
±0.5kV, ±1kV,
±2kV
(common
mode)
During the
test, the
EUT can
operate as
intended
The network power
supply should have the
quality used in a typical
commercial or hospital
environment.
GB/T 17626.11
Votage dips and
interruptions to
AC Power Port
IEC 61000-4-
11:2014
0%UT
0.5T
(0°
45°
90°
135°
180°
225°
270°
and 315°)
0%UT
1T(0°)
70%UT
20T()
0%UT
250T(0°)
0.5T(10ms);
1T (20ms);
25T(500ms)
;
250T(5s)
The network power
supply should have the
quality used in a typical
commercial or hospital
environment. If the A7+
user needs continuous
operation during a
power outage, it is
recommended that the
A7+ be powered by an
uninterruptible power
supply or battery.
Power
Frequency
Table 4 of IEC
60601-1-
2:2014 30A/m,
30A/m Suitable for most
environments, if there is
no industrial magnetic
B ureau Veritas