Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment
The A7+ TouchCare® System is intended for use in the electromagnetic
environment specified below. The customer or user of the A7+
TouchCare® System should assure that it is used in such an
electromagnetic environment.
magnetic fields
IEC 61000-4-8
50HZ and
60HZ
equipment nearby, the
magnetic field strength
will not exceed 400A/m
Proximity fields
from RF
wireless
communication
equipment
IEC 61000-4-
3:2006+A1+A2
Table 9 of IEC
60601-1-
2:2014
During the
test, the
EUT can
operate as
intended
Recommended
separation distance
d = [12/E1] P
80 MHz to 800 MHz
d = [23/E1] P
800 MHz to 6 GHz
Where P is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter
manufacturer and d is
there commended
separation distance
in meters (m).
Field strengths from
fixed RF transmitters, as
determined by
an electromagnetic site
survey, should be less
than the compliance
RF
electromagnetic
field immunity
test
IEC 61000-4-
3:2006+A1+A2
IEC 61000-4-
3:2006+A1+A2
10V/m (for
home
healthcare
environment
and
professional
healthcare
facility
environment)
80 MHz~2.7
GHz
10V/m (for
home
healthcare
environme
nt)
3V/m (for
professiona
l healthcare
facility
environme
nt)
80
MHz~2.7
GHz