Users Manual

227
11 Appendix II: Technical Information
11.1 PDM Specifications
Model: FM-018
Size: 76.2 x 48.4 x 9.375mm
Weight: 42.4 g
Screen: 2.4 in
Operating Temperature Range: +5°C ~ +4C
Operating Relative Humidity Range: 20%~ 90%RH
Operating Atmospheric Pressure: 700~1060 hPa
Storage Temperature Range: -10°C ~ +5C
Storage Relative Humidity Range: 20%~ 90%RH
Storage Atmospheric Pressure: 700~1060 hPa
Classification: Internally powered, Continuous operation
Battery: Built-in 3.8 V polymer lithium ion battery
Power: 5.0VDC, 1.0A
Battery Life: Approximately 1 week once fully charged.
Data Storage: Automatically stores the previous 90 days’ data
Wireless Communication Distance: 10 m with the Transmitter, 4 m with the
insulin pump
Alarm Type: Visual, audible and vibratory
Volume: 52.3 dB(A) measured from 1 m distance
Limited Warranty: 4 years
Dust-proof and Waterproof Rating: IP22
11.2 Patch Pump Specifications
Model:
Pump Base: JN-022
Reservoir Patch: MD-JN-011
Size: 56.5mm x 33.3mm x 13.3 mm
Weight: 21.5 g (without insulin)
Operating Temperature Range: +5 °C ~ +40 °C
Operating Relative Humidity Range: 20%~90%RH
Operating Atmospheric Pressure: 700~1060 hPa
B ureau Veritas