Users Manual

231
11.3 Transmitter Specifications
Model: MD1026
Size: 36.1 mm x 19.4 mm x 7.2 mm
Weight: 3.57 g
Operating Temperature Range: +5°C ~+40°C
Operating Relative Humidity Range: 20%~90%RH
Operating Atmospheric Pressure: 700~1060 hPa
Storage Temperature Range: -10°C~+55°C
Storage Relative Humidity Range: 20%~90%RH
Storage Atmospheric Pressure: 700~1060 hPa
Battery: Built-in 3.7 V polymer lithium ion battery
Waterproof Rating: IPX8 (2.5 m, 60 min)
Classification: Type BF equipment, Continuous operation
Data Storage: Automatically stores the previous 14 days’ data
Wireless Communication Distance: 10 m
Limited Warranty: 1 year
11.4 Glucose Sensor Specifications
Model: MD3026
Storage Temperature Range: +2°C~+3C
Storage Relative Humidity Range: 20%~90%RH
Storage Atmospheric Pressure: 700~1060 hPa
Glucose Range: 2.2~22.2mmol/L (40~400mg/dL)
SterilizationMethod: By radiation
Sensor Life: Up to 14 days
11.5 CGM System Accuracy
A multi-center, randomized clinic study is designed to determine the Sensor
accuracy in adults with Type 1 or Type 2 diabetes. In-clinic testing consisted of
frequent venous blood sample testing (by Yellow Springs Instrument 2300 STAT
Plus™ Glucose Analyzer, YSI) on a random day in the Sensor life. The accuracy
is based on the percentage of CGM glucose readings that are within ± 20%, 30%
or 40% of YSI values at glucose values at or above (>=) 100 mg/dL (5.6