M PAT400 Series Portable Appliance Testers User Manual
Thank you for purchasing the Megger PAT400 portable appliance tester. For your own safety and to get the maximum benefit from your instrument, please ensure that you read and understand the safety warnings and instructions before attempting to use the instrument. These instruments are designed and manufactured by: Megger Instruments Ltd Archcliffe Road Dover Kent CT17 9EN England Megger Instruments Limited reserves the right to change the specification of these instruments at any time without prior notice.
CONTENTS X.1 Unpacking the carton.............................................................................................................................. 5 X.2 Safety Warnings ...................................................................................................................................... 5 X.3 Symbols used on the instrument........................................................................................................... 6 1. Instrument Layout.............................
11. PAT400 enhancements: Firmware release V2.40 ................................................................................... 70 11.1 User names, account status and Login .............................................................................................. 70 11.2 Power Up and Log in/Log out options ................................................................................................ 71 11.3 Managing accounts..........................................................................
=================================================================== X.1 Unpacking the carton Unpack the carton contents carefully. There are important documents that you should read and keep for future reference. Please complete the pre-paid warranty card and return it to Megger Instruments Limited as soon as possible to help us reduce any delays in supporting you should the need assistance.
present due to a potentially faulty appliance Assets should not be routinely flash tested. Where flash testing is required, refer to further guidance on Flash testing, section A7 Replacement fuses must be of the correct rating and type. Refer to section 1.8 The USB connection marked thus should only be used by approved service personnel; nothing should be connected to the USB port during testing. Only use a NiMH rechargeable 8.
1. Instrument Layout 1.1 Overview of the PAT400 The PAT400 series of portable appliance testers are fully automated testers for the fast safety testing of portable electrical equipment. The PAT400s will operate from 230 Vac or, for UK variants 230Vac and 110 Vac supplies. The PAT400’s are ready to start testing, however the warnings must be read and understood before the testers are used.
1.
1.3 Button functions Button functions Power down – Prevents battery discharge Hot keys. Returns to HOME screen Function depends on adjacent display icon. Used for fast selection of options Escape – Exit screen without saving 1.
1.5 Display information Menu bar – Provides navigation path and menu options. See NOTE below Main display area Display and test result information Hot keys Quick access to frequently used test groups, and other options. Additional menu options NOTE: Users tab is only available on Firmware version V2.4 and above. 1.6 Screen navigation Use the UP/DOWN and LEFT/RIGHT arrows to navigate around the menu and asset screens. Use the OK button or CR (enter key) on the keyboard to select or confirm an action.
Free form box. Text or numbers can be entered as required. Text can be entered from a keyboard or barcode scanner. In some instances a “free form” box can become a “drop down” box, such as the asset ID field. In this case it will appear with the drop down icon as below: 2) Drop down boxes: Drop down boxes already contain a list of items. To access the list, navigate to the drop down box using the UP/DOWN arrows then press the OK button. Drop down box. Press OK to select from the available list.
re-started within 5 minutes. The battery shall only be replaced by an equivalent rechargeable battery when the re-boot facility ceases to work. To replace the rechargeable battery: 1. Switch off the instrument. 2. Disconnect the instrument from any electrical circuits. 3. Remove the battery cover. 4. Remove the old battery and refit new battery following correct polarity (as marked on the battery compartment) 5. Replace the battery cover.
For Fuse replacement Warning: 1. 2. 3. 4. 5. 6. 7. Fuses are at mains potential. The electrical supply MUST be disconnected prior to removing the fuse cover. Switch off the instrument. Disconnect the instrument from any electrical circuits. Remove the fuse cover on the underside of the PAT tester. Remove each fuse in turn and check for failure. Replace as necessary. If no fuses have failed, contact Megger Technical services group on 44 (0) 1304 502 102. Replace fuse cover when complete.
2. Getting started 2.1 Switching ON the PAT400 1) Connect the instrument to a suitable electrical supply, either 230 Vac (or 110 Vac for UK instruments). For testing 230 V electrical equipment, connect the PAT tester to a 230 V electrical outlet. For testing 110 V electrical equipment, connect the PAT tester to a 110 V electrical outlet, using the optional 110 V to 230 V supply lead adaptor. Refer to section 3 for further details.
3) Using the RIGHT/LEFT and OK keys to enter the date and time Date format is DD / MM / YYYY Time format is 24 Hr Hr / Min / Sec 4) Press the SAVE HOTKEY to accept changes. Further changes can be made in the menu SETUP options section 10.5 On completion of the initialisation the following warning screen is displayed when the PAT tester is switched on.
2.2 HOME screen explained The HOME screen is the start point of all testing. Menu functions: Home Setup DATA Users 2.
3. Testing with a PAT400 3.1 Connecting an asset to the PAT400 Assets are tested through the test connectors on the top of the instrument. An asset is simply plugged into the appropriate socket. A power cord (IEC lead), or an extension lead needs to be “wrapped back” to the small IEC socket on the top of the instrument. This allows automatic testing of continuity and polarity without connecting the remote probe.
on the PAT400. 3.1.3 Connecting an extension lead To connect an extension lead to the PAT400: 1) Connect the mains plug to the mains test socket on the PAT400 2) Using the extension lead adaptor, connect the furthest socket on the extension lead to the IEC socket on the PAT400. To connect a 110 V power lead, use the 110 V to IEC plug adaptor (optional accessory). 1) Connect the 110 V mains plug to the 110 V test socket 2) Connect the free end of the 110 V power lead to the 110V adaptor plug.
Option 1: Starting a TEST – USING HOT KEYS .1 Select a Class I, Class II, IEC/extension lead or RCD test group by pressing one of the HOT KEY options: A list of TEST GROUPS are displayed that are relevant to that type of equipment Example: Class I test groups Class I test groups Class II test groups Class IEC (Power) and extension lead test groups Plug-In RCD test groups Add Asset - fast access .2 Select one of the test groups by pressing the relevant number, then the TEST button.
If the asset is already in the database the details of the asset will be displayed as below: Press the TEST button and the test sequence will commence. Go to 3.3 TESTING PROGRESS. Alternatively to search through the asset database, press the FIND ASSET down box of Asset IDs will appear. Hotkey. A drop If you know the Asset ID you wish to find, start typing in the Asset ID. The search engine will jump to the Asset IDs that start with those characters. To ADD an asset, simply press the ADD ASSET .
Copy last asset added from list Copy last asset added and increments Asset ID to next available number. * Indicates mandatory fields that are required to save the asset data. Further information on the COPY LAST ASSET ID Hotkey and COPY & INCREMENT ASSET Hotkeys can be found in section 5. (A new asset ID can be added and saved at this stage, but it is recommended that this is added after testing is completed). .4 Press the DOWN arrow to access TEST GROUPS. .
On completion of a test, only Asset IDs and Test groups have to be added before the test is completed. All other data is optional. NOTE: and type in the test If NO Asset ID is entered, press the Test Group Hotkey group. Alternatively press the hotkey twice to select the test group drop down box. Press OK to list all test group options, scroll down the list to select desired test group and press OK to accept. NOTE: Pressing the Right arrow key will “Page Down” the list.
Additional information can also be added, but this is not essential. NOTE: .4 3.3 When scanning an Asset ID that exists in the database under a location that is different to that set in the PATs current status, a yellow warning indicator will show on the Location field as below. Go to section 10 for additional information on Clients and locations. Press the TEST button. Go to section 3.3 TESTING PROGRESS for further information on the test sequence.
Test results screen List of tests in test group Green – Test passed Amber – Test in progress Grey – Test pending Test results column Test limits column Pass limit set for this test Actual measurement result Current test values or Instructions during the test sequence Interruption to automatic testing Testing proceeds automatically unless an action is required from the operator. Follow “on-screen” instructions as they appear. 3.
The insulation test probe is connected to all exposed metal parts to ensure electrical separation from earth. 3.4.1 Test lead resistance null The resistance of the Bond lead can be “nulled” to remove the lead resistance from the measured resistance. This is important when testing very low resistances where the test limit is set to 0.1ohms. Also, the resistance of an extension lead (IEC or Power cord) can be nulled, if the lead is longer than the Red test lead supplied with the PAT.
4) Press the TEST button. The PAT400 will display the value of the test lead as a Null SET value, and display the The 5) symbol in the top of the screen as below: displayed inside means an extension lead null is also set. To remove the Bond Lead Null select the “Remove Lead Null” and press the TEST or OK button. The symbol will disappear. 3.4.2 Extension Lead Adaptor Resistance Null Some tests require using an IEC lead or the short extension lead adaptor supplied with every PAT tester.
2) Press the NULL highlighted: Hotkey. The Lead Null screen will appear with the Set Lead Null option (If an extension lead null value is already set the lead null symbol will be displayed) 3) Connect the extension lead adaptor between the Mains test socket and the IEC socket on the front panel as below: 4) Scroll down to select “Set Extension Lead Null” on the menu 5) Press the TEST button.
Only if the operator is certain the equipment is not a short circuit, press TEST, otherwise press Esc to abort the test 3.6 Aborting a test All tests can be aborted at any time by pressing the TEST button or Esc (escape) button. 3.7 Completion of testing On completion of testing all test fields should turn green as below. The test results are displayed in the centre column with the pass limits for that test group on the right.
Repeat test sequence Repair code Print bar code label & SAVE results to memory SAVE results to memory Example: Asset has failed the insulation test. Pass limit set to 1.0 MΩ, Measurement recorded at < 0.1 MΩ. The same options are available as an asset that has passed. See Section 3.7 for completion of testing. 3.9 Visual failure If the current test group contains a visual test sequence then the visual test popup will appear.
3) Three types of information can be entered: a) Notes - To describe the type of failure. b) Check boxes - Quick entry for standard failure types. c) Repair code - See Section 3.10 below 4) Press the TEST key to complete the test process. The display will now show the TEST FAILED screen. 5) Press the SAVE Hotkey to save the results or the SAVE and PRINT Hotkey to save the results and print a barcode label. 3.10 Repair code A repair code can be entered at the completion of testing.
On completion of the repair code screen press the TEST key to log the changes, then press the SAVE or SAVE and PRINT Hotkeys to store the results. 3.11 Cable compensation A power cord, appliance lead or extension lead will frequently exceed the nominal length and consequently the resistance limit set in the standard test groups.
OR Press Esc to fail the Asset and save or abort the results, OR Press “R” to re-run the continuity test. 2 3) Enter the cable Length in metres and the conductor Cross-Section (mm ) OR a Resistance can be entered if this is known. 4) Press TEST to complete the test. The PAT400 will now proceed to the next test if the resistance was a PASS or to a fail screen if the resistance was still outside the modified limit. See also Section A.6 for cable length limitations and RCD protection on long leads. 3.
The PAT400 will test for a good supply earth during initialisation. A faulty or missing earth will return the error: Error 73 – Mains supply Earth Fault 3.12.4 Parallel earth paths If the PAT400 detects more than one earth path, the warning: Unable to verify earth connection, Parallel paths may exist. Refer to user guide. The PAT400 has detected more than one earth path during the Earth continuity, Earth bond, Insulation or Earth leakage tests. Refer to Appendix A.8 for further details. 3.12.
4. Using the memory 4.1 Important overview All records are stored under a “client”. The PAT400 can store up to 100 (50 on firmware releases prior to 2.20) clients. Each client can have up to 2 000 (50 on firmware releases prior to 2.20) Locations. There is a maximum of 25 000 total locations per database. The client and locations are configured in SETUP, refer to Section 10. The PAT400s are shipped with the client and first location set to “default”.
Definitions of fields Field Limit Description Asset ID:* 10 000 Unique number or combined letters and numbers. Enter by typing from the keyboard or scanning a barcode. Separate clients can use the same ID. But one Client cannot have two identical Asset IDs. Test Group:* 100 Drop down box of available test groups. Press OK to access list, DOWN/UP arrow to select and OK to enter. Description: --- General description of asset (Cooling fan, hand drill etc).
3) Enter the Asset ID. This can be via the keyboard or scanned in by barcode scanner. If an Asset ID already exists in memory, a “duplicate Asset ID” warning will be displayed. In this case, enter a different Asset ID number. Examples of Asset ID formats: (Asset IDs are limited to 20 characters) 0001, 0002, 0179, 7082 MEGGER0001, MEGGER0002 AA0001,AA0002, BF0001, BF0002 4) Press OK 5) Press the DOWN arrow and complete the additional asset data as required.
5. Adding assets to memory Assets can be saved to memory after they are tested, as above, or before any testing starts. Copying a previously added asset using the COPY ASSET INCREMENT ASSET ID Alternatively: 5.1 Hotkey or COPY ASSET AND Hotkey in Section 5.2. Asset data can be loaded from PowerSuite via a USB stick, Section 7.2. Adding assets to memory before testing 1) From the HOME screen, press the ADD ASSET appear: Hotkey.
5.2 Copying the previously added asset 1) From the HOME screen, press the ADD ASSET appear: 2) To recall the last asset to the screen press the COPY ASSET Hotkey. This will recall the last asset added to the database of the current Client. The Asset ID can then be edited as required. 3) Change the Asset ID. The SAVE changed. Hotkey.
4) Change any other asset information as necessary. 5) Press the SAVE Hotkey to save the ASSET to memory.
6. Editing and deleting assets and results All asset data can be edited or deleted as required. For deletion of Test Groups see section 8. For deletion of Client and Locations see section 10. 6.1 Editing asset data 1) From the HOME screen, enter the SETUP menu option by pressing the RIGHT arrow key. EDIT asset 2) Press the EDIT ASSET Hotkey. The EDIT ASSET screen should be displayed. Enter an Asset ID to be edited. Alternatively, to list the assets in the database, press the FIND ASSET Hotkey.
closest match as you type. Note: The first 2 000 Asset IDs are loaded into the drop down list. To access the next 2 000 assets, press and hold the RIGHT ARROW to scroll down to the bottom of the first 2 000 assets. Select ‘Next…’ from the bottom of the list. The Asset ID list will now load the next 2 000 assets. OR Where available type ‘next…’ to jump to the end of the list or ‘previous…’ to jump to the start of the list. 3) Edit the asset details as required and press the SAVE key to store changes.
7. DATA – Test results and data transfer Data storage – backup and restore The PAT400 can store up to 10 000 electrical assets with associated test records. Megger STRONGLY RECOMMEND this data is backed up frequently. MEGGER CANNOT ACCEPT RESPONSIBILITY OR ANY LOST DATA HOWEVER CAUSED. The backup and recall operations are extremely fast. 10 000 records can be backed up to a USB stick in <20 seconds. Smaller databases are significantly faster. The file format for data backup is a db file.
5) Press OK to initiate backup. If the file name clears without backup starting, the file name contains excluded characters. Only 0 to 9 and A to Z should be used. 6) On completion the display will show the message “Backup complete”. There are no limits to the number of times you can backup a database. The stored file is a “.db” file. Do not attempt to edit this file as this can corrupt the data and prevent further use of that file.
WARNING: Importing or restoring a file will overwrite the existing assets, results, test groups, clients and locations in the database. 3) Press OK to continue. 4) The PAT400 will search for any database files in the root directory. A drop down box appears. Press the OK key and then the DOWN arrow to scroll through the list, or use the RIGHT arrow key to page through the list. OR Type the desired file name to jump to the matching item in the list.
6) Press OK to accept the file for import/restore. 7) The file will be imported. On completion a message will be displayed and the buzzer will sound. DO NOT SWITCH OFF THE PAT400 DURING THE USB OPERATION! 7.3 WARNING: Importing or restoring a file will overwrite the existing assets, results, test groups, clients and locations in the database. Always ensure existing data has been backed up prior to importing new data.
2) Select Export Results to USB and press OK. The following screen will appear. 3) Enter a name for the CSV file. Up to 8 characters can be used. Use letters and/or numbers. Press OK when ready. 4) The CSV file will be saved to the USB memory stick. On completion the following message will be displayed and the buzzer will sound. Note: Exporting large amounts of data to CSV can take a significant time to complete. Once started a CSV export cannot be aborted. 7.
The ‘View Test Results’ screen will be displayed as below: FIND asset ADD asset EDIT 3) Enter the Asset ID or press the FIND ASSET this may take a few seconds. Hotkey. If there are a large number of Assets IDs, 4) When the desired Asset ID has been found, Press OK to accept. The VIEW TEST RESULTS Hotkey will appear as below.
5) Press the VIEW TEST RESULTS asset as below. 6) To exit this screen press HOME or Esc. Note: 7.4.2 Hotkey. The Test Results screen will be displayed for that Barcodes can be printed by using the PRINT BARCODE Hotkey Printing barcode labels Duplicate barcode labels can be printed from the VIEW TEST RESULTS page. This allows replacement of labels if they are damaged or multiple labels are required for an asset. To print a barcode label: 1) Enter the Test Result screen as above.
8. Test groups Overview of test groups Test groups are used to simplify the testing of an asset. A test group is a collection of tests that when run, perform all the tests allocated to that test group. The PAT400’s are shipped with a range of test groups for the majority of electrical equipment that will require testing, as detailed below. Additional test groups can be added up to a limit of 100. Existing test groups can be edited or deleted as required. 8.
Note: If the number of test groups reaches 100, the ADD 4) Hotkey will change to a Icon. Complete the fields as required. An example of a typical standard Class I test may look like: Next page Definitions to fields: 1 NAME : 1 DESCRIPTION : Test group name - This can be up to 10 characters - For compatibility with PAT4 see section 11 Text string - Keyboard or Barcode entry SUPPLY: Asset operating voltage.
Options FLASH - Class 1 - Class 2 - EXT Check box – configuration options available on second page LEAKAGE Check box – configuration options available on second page LOAD Check box – configuration options available on second page RCD Check box – configuration options available on second page 1 5) These fields require mandatory data. They cannot be left blank. When complete, press the NEXT PAGE will appear as below: Hotkey.
Class I Class II Power cord / Extension lead RCD Add Asset Using the DOWN ARROW, navigate to the desired row (1 – 9) as below: Press OK to access the drop down list of available test groups. 3) Select the required TEST GROUP and press OK. The TEST GROUP will be allocated to the selected field.
NOTE: 8.3 The TEST GROUPS can be freely allocated to any of the nine fields and in any of the four Hotkeys. If desired a CLASS I TEST GROUP can be allocated to the RCD Hotkey for example. Editing a test group Each test group can be edited or deleted NOTE: When restoring a database the test groups will also be overwritten. Always backup your test records before restoring a database. To edit a test group: 1) From the HOME screen use the RIGHT ARROW key to select the SETUP tab.
Select the test group to be edited and press OK to accept. The test group settings can be adjusted and saved as per the test group ADD function above. WARNING: 8.4 Changing a Test Group will affect all assets that use that test group. Deleting a test group To delete a test group: 1) From the HOME screen use the RIGHT ARROW key to select the SETUP tab. 2) Press the TEST GROUP MANAGER 3) Press the DELETE Hotkey. Hotkey. Delete 4) Press OK to activate the drop down list.
Note: 8.5 A test group cannot be deleted whilst there are assets in the database that are using that test group. Repeated Tests It is possible to create test groups with repeated Bond, Continuity, Insulation and Flash tests. An example screen is shown below. Repeated tests are setup by choosing an option in the Cycles column. Bond and Continuity test have the ability to be repeated a set number of times (1 to 10) or repeated indefinably by choosing the ∞ symbol.
9. Quick test - QT The Quick Test (QT) key provides instant access to continuity, bond, insulation, leakage, RCD, load (Operational) and flash testing, without having to create a test group. NOTE: Some tests shown below may not be available, depending on the instrument being used. Running a quick test 1) Press the QT key on the keyboard. The display will show a numerical list of options. Example of QT screen options for a PAT450 2) Press the number key associated with the test you wish to run.
5) Follow on-screen instruction as necessary. On completion of the test the final measurement will be displayed. Quick test times can be changed in SETUP. Refer to section 10.
10. SETUP menu options The SETUP menu options allow changes to the instrument configuration that cannot be accessed by Hotkeys. 10.1 Clients The PAT400s are shipped with the CLIENT and LOCATION set to DEFAULT. This can be changed, or additional Clients and locations can be added, edited or deleted. The DEFAULT CLIENT and LOCATION can only be deleted after other additional CLIENTS AND LOCATIONS are added. 10.1.1 Adding Clients 1) Select CLIENT from the SETUP menu and press OK.
5) On completion, press the SAVE Hotkey. 6) The display automatically jumps to the ADD LOCATION screen. 7) Add the location details. Only Location name is mandatory. 8) On completion, press the SAVE Hotkey. A new CLIENT and LOCATION has been added to the database. The current CLIENT and LOCATION has been updated to the new CLIENT and LOCATION automatically. 10.1.2 Editing and Deleting Clients This function should be used to change the details of a Client.
3) Press the DELETE 4) A popup message will ask you to confirm. hotkey. 10.1.3 Changing Clients To change between different Clients in the PAT400: Hotkey. 1) From the SETUP tab press the SELECT CLIENT 2) Press the OK key to open the drop down list. Select the desired Client and press OK. 3) Press the SAVE 4) The display will now change to the SELECT LOCATION screen. Select the appropriate Location from the drop down list. 5) Press OK to accept the new location.
10.2 Locations Multiple locations can be added to one CLIENT, up to a maximum of 2 000*. To add a LOCATION, select the LOCATION option from the SETUP menu. Follow the same process as adding Clients in Section 10.1. * Firmware revisions earlier than 2.2 were limited to 50 locations per Client. 10.2.1 Adding Locations 1) Select LOCATION from the SETUP menu and press OK.
1) Select LOCATION from the SETUP menu and press OK. In the Location screen select the EDIT Hotkey. 2) Select the desired Location then press the NEXT PAGE 3) Change the Location details as required. It is not possible to change the associated Client. 4) After the details have been changed press the SAVE Hotkey. hotkey. Delete Location: Warning: 1) Deleting a Location will also delete all assets and test results associated with that Location Select LOCATION from the SETUP menu and press OK.
2) Select the desired Location then press the NEXT PAGE 3) Press the DELETE 4) A popup message will ask you to confirm. Hotkey. Hotkey. 10.2.3 Changing Locations You may only change to a Location in the current client. 1) In the SETUP tab press the SELECT LOCATION 2) Select the desired Location and press OK. Hotkey.
3) 10.3 Press the SAVE Hotkey to store the changes. Company The Company page provides data entry for the testing company to add their company details to the PAT400. The “Label Text” is a field that is used to add text to printed labels. For example you may wish to add the testers name to a label. This field appears at the top of the barcode label. 10.4 1) From the HOME screen, enter the SETUP menu option by pressing the RIGHT arrow key. 2) Select Company from the menu options and press OK.
Insulation test Bond test Load (Operational) test Leakage test To adjust the QUICK TEST time settings: 10.5 1) Select QUICK TEST KEYS from the SETUP menu options. 2) Press OK to display the Quick Test Config screen. 3) Navigate to the desired field to adjust the test time. 4) Press ESC to save and exit. Date / Time settings To change the Time or Date settings: 1) From the HOME screen, enter the SETUP menu option by pressing the RIGHT arrow key.
3) Press OK to accept. 4) For DATE settings, use LEFT/RIGHT arrow keys to select the field to be changed. Enter the desired values using the keyboard. For TIME settings, use the DOWN arrow key to select TIME, then the LEFT/RIGHT arrow keys to select the fields to be changed. 5) 10.6 Press the save Hotkey to accept changes.
10.7 2) Press the OK key and the language selection screen will be displayed. 3) Use the UP/DOWN arrow keys to select the required language. 4) Press OK to save. Restoring default settings WARNING: Restoring factory default settings will overwrite all stored data and settings. Stored data will be lost. Megger strongly recommend backing up all data prior to restoring factory settings. Backups can be restored if necessary.
Existing records and test groups should be backed up prior to restoring factory settings. 2) Press OK to access the Restore Defaults screen. 3) Press OK to proceed, or Esc to abort. 4) A popup message will ask you to press Y to confirm restore to defaults. 5) The PAT400 will then reboot with the default settings.
10.8 ABOUT - PAT400 software version and Megger contact details To select the product information and Megger contact details: 1) Using the RIGHT ARROW key, select SETUP in the top menu option, then the DOWN arrow to select ABOUT option. 2) Press OK to access the ABOUT screen. 3) To exit the ABOUT screen press Esc.
11. PAT400 enhancements: Firmware release V2.40 The PAT400 now includes the following additional features: Multiple operator accounts Login (PIN) facilities Security levels (Supervisor/User accounts) Experience levels (Expert mode) Reverse polarity earth leakage testing The additional functionality is located under the new “Users tab” in the menu options as below: 11.1 User names, account status and Login The PAT400 supports single or multiple operators.
Username: This can be changed as required Additional users can be added Maximum field length - 30 User status: Indicates the level of control and configuration available to the operator 11.2 Power Up and Log in/Log out options Factory default Power-up When powering up the PAT400 for the first time there is no need to log in. Unless a PIN is added or additional accounts created, the Login/Logout option is not enabled.
3) The PAT400 will display the warnings screen. If the wrong PIN is entered, the display will show the warning: “Incorrect PIN entered” Retry with the correct PIN. 4) Successful log in will then display the “Warnings” screen. Press OK to indicate you have read and understood the warnings. 5) The PAT400 will display the HOME screen. Power-up and login with multiple accounts To add a PIN to the “default” account, see section 11.3.3 Using PIN numbers.
3) Use the UP/DOWN arrows to select the required User name and press OK. 4) Enter the PIN for the new “Username” and 5) Press the Hotkey. Logging Out & Power down Unless the operator has logged OUT of their account on Power-down, the operator will remain logged IN to the PAT400 on Power-up. To log out of an account either: 1) Press the RED power OFF button.
a) Supervisor status – Supervisor box checked An account with “Supervisor” status has unlimited access to all functionality and configuration of the PAT400. Any account with “Supervisor” status can create additional accounts and control the status of other accounts, including deleting other accounts. b) User status – Supervisor box unchecked An account with “User” status has limited access to functionality.
3) Press the SAVE Hotkey to save the change. If the Feature is disabled, the Expert mode check box will be grayed out. Reverse L-N Polarity (Only available in some models) Refer to section 12 Reverse L-N Polarity. 11.3.2 Changing the default Supervisor account The factory set user account name is called “default”. This can be changed if required, see section below To change the default account name: 1) From the HOME screen use the RIGHT arrow key to select the “Users” tab.
Usernames must be between 3 and 30 characters, A-Z and 0-9 (upper or lower case). 5) When completed press OK. 6) To add a PIN, follow section 11.3.3 To add a PIN to an account step (3), OR If no PIN is to be entered, press the SAVE Hotkey to save the change. 11.3.3 Using PIN numbers PIN numbers on accounts are optional. All test results are tagged and saved with the name of the account holder (the user name).
2) Press the Edit Hotkey 3) Use the DOWN arrow to select the PIN field. 4) Enter a PIN number from 0000 to 9999. Only use numbers from 0 to 9. 5) Press the DOWN arrow to select the “Confirm PIN” field. 6) Enter the same PIN number. 7) Press the SAVE Hotkey to store the changes. Once logged out, the operator will need to log into that account by entering a valid PIN.
2) Press the Edit Hotkey. 3) Use the DOWN arrow to select the PIN field. 4) Enter a new PIN number. 5) Press the DOWN arrow to select the “Confirm PIN” field. 6) Enter the same (new) PIN number. 7) Press the SAVE Hotkey to store the changes. If the account has “User” status: 1) From the HOME screen use the RIGHT arrow key to select the USERS tab and select the “Change PIN” option.
2) Enter a new PIN and press the DOWN arrow to the Confirm PIN field. 3) Enter the Same PIN. 4) Press the SAVE Hotkey to store the changes. To remove a PIN from an account: If you account has “Supervisor” status: 1) Use the same method as to change a PIN, but delete the PIN number entirely. 2) Repeat for the Confirm PIN field. 3) Press the SAVE Hotkey to store the changes.
2) Using the arrow keys, navigate to “Manage Users” option in the USERS Tab and Press OK. Note: An additional Delete 3) Press the ADD hotkey may be present if there is already more than one account. Hotkey to bring up the Add User screen.
4) Enter a user name. Names must be between 3 and 30 characters, A-Z and 0-9 (upper or lower case). 5) Press the DOWN arrow. Enter a PIN if required. The PIN must be 4 digits using only numbers 0-9. 6) Press the DOWN arrow. Enter the same PIN in the Confirm PIN field. 7) Press the DOWN arrow to Expert Options. Press OK to enable this option if required. 8) Press the Down arrow to Supervisor. Press OK if this account is to have Supervisor status. 9) Press the DOWN arrow to Reverse L-N Polarity.
3) If this is the account to be deleted, press the Delete Hotkey. Or Press the OK key to display a list of accounts and use the DOWN/UP arrow keys to highlight the require account and press OK. Hotkey to delete the account. 4) Press the Delete 5) The Warning message will be displayed. Press OK to confirm or Esc to abort. Notes: Deleting an account whilst logged into the same account will log the operator out of the PAT tester. 11.
Pressing OK will remove all users from the PAT400 and create a single Supervisor account. If necessary, contact your supervisor before continuing, or press ESC to abort. 4) Press OK to continue. 5) The display will show a standard warning screen. 6) Press OK after reading and understanding the warnings given. The PAT400 is now logged into a standard Default user account as a “Supervisor”.
12. Reverse L-N Polarity (not available on all instruments) In countries where the power plug can be reversed in the socket (example: Schuko) the earth leakage tests can be configured to test in normal “forward” polarity or “normal and reversed” polarity. Checking the “Reverse L-N Polarity” box in the “Add User”, “Edit User” or “Expert Options” screen enables the forward and reverse earth leakage testing. Reversing the test is performed automatically and the worse of the two results are recorded.
5) Press the SAVE Hotkey to store the changes. Operators who have “Supervisor” status or “Expert” user status may also configure the Reverse L-N Polarity feature in the Expert Options screen. 1) Use the UP or DOWN arrow key to highlight the Reverse L-N Polarity check box. 2) Press the OK key to enable or disable the reverse L-N testing.
3) Press the SAVE Hotkey to store the changes.
13. Measure PE Option (not available on all instruments) Some PAT testing practices allows the Protective Earth bond testing to be skipped if it is deem impossible to access an exposed PE contact area. For PAT testers configured to allow for such situation there is a “Measure PE Option” checkbox made available in the Expert Options screen. Changing the existing state of the “Measure PE Option” checkbox will cause the Save appear. User must press the hotkey to save the new status of the checkbox.
14. PAT4 compatibility The PAT400’s have additional test capability that does not exist on previous PAT testers. To maintain compatibility when transferring data between the PAT 4 and the PAT400s, the following should be considered: Test options The PAT400s include additional tests not available on the PAT4.
To ensure hardware and software compatibility, refer to the Megger website for latest PowerSuite software and PAT400 firmware revisions. www.megger.
15. Care and maintenance The PAT400 should only be opened or repaired by an approved Megger service centre or by Megger Instruments Limited. To clean the instrument; disconnect the instrument from the power supply then use a clean cloth dampened with water or isopropyl alcohol. To clean the display window only use a lint free cloth. For warranty repairs see appendix E.
16. Carry case The carry case for the appliance tester has a lead storage pouch in the lid of the case when opened. This is designed for basic lead and document storage. Further items can be stored in the pouch. If it becomes difficult to close the case, the storage pouch can be removed from inside the case and attached to the front using the straps on the reverse of the pouch.
Appendix A – Test types explained Each test group is constructed from a series of individual tests. The tests are selected depending on the electrical construction of the asset under test. The operator must understand the type of electrical construction of the asset before a correct selection of tests can be made. The tests available on the PAT400’s are as follows: A.
also fail due to increased resistance induced by long cable lengths. When a bond test fails the PAT400 will display a lead compensation table, allowing the pass limit to be modified. See section 3.11 and Appendix A.6 for further details on lead compensation. A.2 Insulation testing This is used to confirm there is separation between the live conductors (live and neutral) and any accessible conductive parts.
Also called absence of potential Test voltage: 230 Vac 50 Hz 9.99 mA 110 Vac 50 Hz 9.99 mA During this test the appliance under test is powered from its normal mains supply and the test lead and probe is connected to the PAT terminal. This probe is applied to any conductive parts on the appliance under test. This probe performs a current measurement with respect to earth. This test can also be performed on three phase appliances powered from their own supply. Warning: A.
A.6 Power lead/extension lead testing Power cords and extension leads are tested in a similar way. The principle differences are the need to use an extension lead adaptor when testing extension leads, as the lead must be looped back to the PAT400 to allow bond and polarity testing. Extension leads or power leads with surge protection should be tested with an Insulation test voltage of 250 V. This prevents the surge protection from operating and artificially failing the lead.
the equipment touching other earthed equipment or from touching the ground. Parallel earth paths affect measurement accuracy and in worst case situations can lead to equipment being passed as safe when there is no proper earth connection through the electrical outlet. b. A hardware fault in the PAT400 c. A combination of the above. Any reports of parallel paths should be investigated. Check to ensure If the BOND (continuity) lead is not connected to the null post on the PAT400.
A.9 PAT400 Factory Default Test Groups Test Groups for UK PAT Variants: Test Group Class Description SC1 SC1-RCD SC1B SC1B-RCD SC2 TO1 TO1F 1 1 1 1 2 1 1 Test Applied Pass Limit Test Duration Used in Standard Class 1 Earthed Equipment Visual Inspection Earth Bond at 10A Insulation test at 500V Operation test (230V) Differential Earth Leakage (230V) ≤0.1Ω ≥1MΩ ≤3000VA ≤3.
TO2 TO2F FAN PC DRY KET KETB EL1 EL1B 110V Tools Double Insulated Visual Inspection Insulation test at 500V Operation test (110V) Touch Current (110V) ≥2MΩ ≤1800VA ≤0.25mA 5s 5s 5s UK only PAT420 PAT450 110V Tools Class 2 + Flash Double Insulated Visual Inspection Insulation test at 500V Operation test (110V) Touch Current (110V) Flash test at 3000V ≥2MΩ ≤1800VA ≤0.25mA ≤0.
EL2 EL3 EL4 EL4B RCD RCD FULL PSC1 230V Extension Leads 2 Core Visual Inspection Polarity Insulation test at 500V ≥2MΩ 5s PAT410 PAT420 PAT450 110V Extension Leads 3 Core Visual Inspection Polarity Earth Bond at 25A Insulation test at 500V ≤0.1Ω ≥1MΩ 5s 5s UK only PAT420 PAT450 EXT 4 way 230V Extension Leads 3 Core Visual Inspection Polarity Earth Bond at 25A (x4) Insulation test at 500V ≤0.
Test Groups for Australian/New Zealand PAT Variants: Test Group Class Description Test Applied Pass Limit Test Duration Used in 1 Standard Class 1 Earthed Equipment Visual Inspection Earth Continuity (200mA) Insulation test at 500V ≤1Ω ≥1MΩ 5s 5s PAT410 PAT420 PAT450 AU-SC1-250V 1 Standard Class 1 Earthed Equipment Visual Inspection Earth Continuity (200mA) Insulation test at 250V ≤1Ω ≥1MΩ 5s 5s PAT410 PAT420 PAT450 AU-SC1-RCD 1 Standard Class 1 Earthed Equipment with RCD Protection Visua
AU-EL4 AU-RCD EXT 1 AU-RCD FULL 1 4 way 230V Extension Leads 3 Core 200mA Continuity Test Visual Inspection Polarity Earth Continuity (x4) Insulation test at 500V Portable RCD 30mA Visual Inspection RCD 30mA test PAT410 PAT420 PAT450 Portable RCD (5 tests) Visual Inspection RCD full test sequence PAT410 PAT420 PAT450 ≤1Ω ≥1MΩ 5s 5s PAT410 PAT420 PAT450 102
Test Groups for International/European PAT Variants: Test Group Class Description Test Applied Pass Limit Test Duration Used in ≥2MΩ ≤0.5mA 5s 5s PAT410 PAT450 N-KL2STD 2 Klasse 2 Visual Inspection Insulation test at 500V Substitute Earth Leakage N-KL2ELEC 2 Klasse 2 elektronisch regeling Visual Inspection Differential Earth Leakage ≤0.5mA 5s PAT410 PAT450 Klasse 1 Visual Inspection Earth Continuity Insulation test at 500V Substitute Earth Leakage ≤0.
E-SC1 E-SC1-RCD E-SC1B E-SC1B-RCD E-SC2 E-FAN E-PC E-DRY 1 1 1 1 2 2 1 2 Standard Class 1 Earthed Equipment Visual Inspection Earth Bond at 10A Insulation test at 500V Operation test (230V) Differential Earth Leakage (230V) ≤0.1Ω ≥1MΩ ≤3000VA ≤3.5mA 5s 5s 5s 5s PAT450 Standard Class 1 Earthed Equipment with RCD Protection Visual Inspection RCD 30mA test Earth Bond at 10A Operation test (230V) Differential Earth Leakage (230V) ≤0.1Ω ≤3000VA ≤3.
Kettles Earthed Visual Inspection Earth Bond at 25A Insulation test at 500V Operation test (230V) Differential Earth Leakage (230V) ≤0.1Ω ≥1MΩ ≤2500VA ≤3.5mA 5s 5s 5s 5s PAT450 Kettles Earthed 200mA Continuity Test Visual Inspection Earth Continuity Insulation test at 500V Operation test (230V) Differential Earth Leakage (230V) ≤0.1Ω ≥1MΩ ≤2500VA ≤3.5mA 5s 5s 5s 5s PAT410 PAT450 EXT 230V Extension Leads 3 Core Visual Inspection Polarity Earth Bond at 25A Insulation test at 500V ≤0.
Appendix B – Barcode sample table The following sheet is an example of a bar code sheet that can be used in conjunction with the optional OPTICON barcode scanner. The sheet allows fast input of standard data, which would otherwise need to be manually typed in or selected from a drop down list.
UK Test Groups Australian/New Zealand Test Groups 107
International/European Test Groups 108
Appendix C - Specification SPECIFICATIONS Electrical supply range PAT-UK 230 V ±10% ±1 V @ 50 Hz ±1% ±0,1 Hz PAT-UK 110 V ±10% ±1 V @ 50 Hz ±1% ±0.1 Hz PAT-NL-ES 230 V ±10% ±1 V @ 50 Hz ±1% ±0,1 Hz PAT-DE 230 V ±10% ±1 V @ 50 Hz ±1% ±0,1 Hz PAT-CH 230 V ±10% ±1 V @ 50 Hz ±1% ±0,1 Hz PAT-PL 230 V ±10% ±1 V @ 50 Hz ±1% ±0,1 Hz PAT-AU 230 V ±10% ±1 V @ 50 Hz ±1% ±0.
Resolution: Display range: Reading corrected to 230 Vac or 110 Vac Extension lead test Tests performed: Polarity test voltage: Polarity checks: ±5% ±100 digits (1000 VA - 3700 VA) 1 VA (0 to 3700 VA) 0 to 3990 VA Bond, insulation & polarity 12 V dc Lead OK Live neutral S/C Live neutral reversed Live/neutral O/C Flash test Flash test voltage: Flash test current: Flash test breakdown current accuracy: Flash test breakdown current resol
Appendix D - Ordering information ORDERING INFORMATION Order Code Included accessories (relevant to territory) Continuity / earth bond lead + probe (black) Extension lead adaptor 13 A BS Extension lead adaptor 110 V BS Extension lead adaptor (ELA) 230 V AU (AS/NZ253112) Extension lead adaptor (ELA) 230V SC (CEE7/7) Extension lead adaptor (ELA) 120V US Flash lead 3.0 kV 3.
Appendix E - Repair and Warranty The instrument contains static sensitive devices, and care must be taken in handling the printed circuit board. If an instrument’s protection has been impaired it should not be used, but sent for repair by suitably trained and qualified personnel.
M Megger Limited Archcliffe Road Dover Kent, CT17 9EN England Tel: +44 (0) 1304 502100 Fax: +44 (0) 1304 207342 Megger 4271 Bronze Way Dallas TX 75237-1017 U.S.A. Tel: +1 (800) 723-2861 (U.S.A. only) Tel: +1 (214) 330-3203 (International) Fax: +1 (214) 337-3038 Megger Valley Forge Corporate Centre 2621 Van Buren Avenue Norristown, PA 19403, USA Tel: +1 (610) 676-8500 Fax: +1 (610) 676-8610 Megger SARL Z.A. Du Buisson de la Couldre 23 rue Eugène Henaff 78190 TRAPPES France Tel : +33 (1) 30.16.08.