9. Quick test - QT
The Quick Test (QT) key provides instant access to continuity, bond, insulation, leakage, RCD, load
(Operational) and flash testing, without having to create a test group.
NOTE: Some tests shown below may not be available, depending on the instrument being
used.
Running a quick test
1) Press the QT key on the keyboard. The display will show a numerical list of options.
Example of QT screen options for a PAT450
2) Press the number key associated with the test you wish to run.
Example - Earth Leakage
Note: Where a test has several options, such as Earth Leakage, a sub menu of options will
appear.
3) Navigate through the sub-menu options if available.
4) Press OK or TEST key to start testing.