5.0 Maintenance 
WARNING: Lethal voltages are present within the chassis of the power amplifier 
whenever AC power is connected. Users should not attempt repair operations on 
components of the power amplifier. 
A great deal of care has gone into every power amplifier to verify that it meets or exceeds all its 
specifications before it is shipped to you. If you are certain the problem lies in the power 
amplifier, check first to determine if the power indicator is illuminated. If it is not, the main AC 
mains breaker has likely tripped. Reset the breaker. 
After resetting the AC mains breaker, the amplifier may immediately cause another breaker trip. If 
this should occur, there could be an amplifier power supply problem and the power amplifier 
requires service.  
CAUTION: To prevent damage to equipment, NEVER replace a fuse with anything 
other than the type specified. Using types other than those specified will void the 
warranty and may cause fire or injury to personnel.  
The output fuse (located on the front panel) should only be replaced with AGC or 3AG types as 
supplied with your system. When removing this fuse, ensure that power is not applied to the 
power amplifier.  
The output fuse, because of its construction, may interrupt the circuit before a circuit breaker on a 
matching network opens. DO NOT substitute a higher value or a different type of fuse. Decrease 
the matching network output voltage tap or decrease the level of the signal generator. Consult 
Meggitt Sensing Systems for more information about inductive tuning or other special application 
assistance. 
For piezoelectric shaker systems employing matching networks, there may be another current-
limiting device (such as a fuse, circuit breaker, or thermal cutout) to the amplifier’s output within 
the matching network. Consult the matching network’s manual for details. 
CAUTION: The power amplifier contains no user serviceable components. Service or 
repair of the power amplifier should not be attempted by anyone other than Meggitt 
Sensing Systems employees or designated representatives. 